Open original FDA record ↗Recall record
- FDA identifier
- Z-2452-2015
- Company in FDA record
- Trilliant Surgical Ltd.
- FDA classification date
- Aug 24, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The cannulation of the 3.0/4.0 Cannulated Drill Bit inner diamter is too small to allow for the guide wire to pass completely through.
Source facts
- Action
- Trilliant Surgical sent an Important Medical Device Advisory Notice to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Customers were instructed they have the option to exchange the affected drill bits at no charge. To initiate the exchange, customers were instructed to contact their local Trilliant Surgical Sales Representative or contact the Customer Service Department directly at 1-800-495-2919 or via email at orders@trilliantsurgical.com.
- Initiated
- 2015-07-20
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-01-26
- Distribution
- Nationwide Distribution including NY, TX, OH, IN, FL, NH, MN, SC, OK, NJ, CA, NC, MO, GA, WI, IA, TN, WA, and Puerto Rico.
- Recall number
- Z-2452-2015
- Recall status
- Terminated
- CDRH record ID
- 139141
- FDA classified
- 2015-08-24
- Recall event number
- 71812