Skip to content

Public FDA record

Z-2452-2018: PROLENE SUTURE 30"(75CM) 4-0 BLUE, D7748

Recall record Z-2452-2018 for Ethicon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2452-2018
Company in FDA record
Ethicon, Inc.
FDA record creation date
Jul 17, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The needle associated with certain suture product codes has the potential to perforate the packaging due to a packaging design issue. A perforated package has the potential to cause a soft tissue injury to a healthcare professional handling the subject product or a patient infection due to a breach of sterility within the packaging if it goes unnoticed by the healthcare professional.

Source facts

Action
On May 24, 2018, the firm distributed Urgent: Medical Device Recall letters to affected U.S. customers. Customers were instructed to do the following: 1. Examine your inventory immediately to determine if you have product subject to this recall (removal) on hand and quarantine such product(s). 2. Remove the product subject to this recall (removal) and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall (removal) has been forwarded to another facility, contact that facility to arrange return. 4. Complete the Business Reply Form (BRF) confirming receipt of this notice and fax or email it to Stericycle at 1-888-473-8015 or Ethicon3681@stericycle.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall (removal). 5. Keep this notice visibly posted for awareness until all product subject to this recall (removal) has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall (removal) and keep a copy for your records. 6. Customers are required to return all unused D-Special PROLENE Polypropylene Suture & D-Special PRONOVA Poly (Hexafluoropropylene - VDF) Suture products that are in their inventory immediately. Only D-Special PROLENE Polypropylene Suture & D-Special PRONOVA Poly (Hexafluoropropylene - VDF) Suture product subject to this recall (removal) returned by September 21, 2018 will be eligible for credit reimbursement. Any product subject to this recall (removal) returned after September 21, 2018 will not be eligible for credit reimbursement. 7. To return product subject to this recall (removal), photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for
Initiated
2018-05-07
Root cause
Package design/selection
Terminated
2020-06-08
Distribution
Accounts in AZ, CA, CO, FL, GA, IL, IN, KS, MA, MN, MO, NC, NE, NJ, NY, OH, OR, SC, SD, TN, TX, VA, WA, WI, WV.
Recall number
Z-2452-2018
Recall status
Terminated
CDRH record ID
165109
FDA record created
2018-07-17
Recall event number
80306

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗