Open original FDA record ↗Recall record
- FDA identifier
- Z-2452-2025
- Company in FDA record
- MEDLINE INDUSTRIES, LP - Northfield
- FDA classification date
- Aug 27, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Affected kits contain recalled Stryker Color Cuff Sterile Disposable Tourniquet Cuffs. Stryker Instruments initiated this recall due to a potential manufacturing defect in which the flange may detach from the bladder at the welded connection point. This detachment compromises the cuff's ability to maintain adequate air pressure, potentially resulting in tourniquet failure and increased risk of patient hemorrhaging.
Source facts
- Action
- On July 8, 2025, the firm communicated the recall to customers who had received affected kits. Medline instructed customers to request stickers to over-label the affected convenience kits with a warning label, indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
- Initiated
- 2025-07-08
- Root cause
- Unknown/Undetermined by firm
- Distribution
- US Nationwide distribution tin the states of CA, FL, IA, and TX.
- Recall number
- Z-2452-2025
- Recall status
- Open, Classified
- CDRH record ID
- 215404
- FDA classified
- 2025-08-27
- Recall event number