Skip to content

Public FDA record

Z-2452-2025: The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care prof

Recall record Z-2452-2025 for MEDLINE INDUSTRIES, LP - Northfield. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2452-2025
Company in FDA record
MEDLINE INDUSTRIES, LP - Northfield
FDA classification date
Aug 27, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected kits contain recalled Stryker Color Cuff Sterile Disposable Tourniquet Cuffs. Stryker Instruments initiated this recall due to a potential manufacturing defect in which the flange may detach from the bladder at the welded connection point. This detachment compromises the cuff's ability to maintain adequate air pressure, potentially resulting in tourniquet failure and increased risk of patient hemorrhaging.

Source facts

Action
On July 8, 2025, the firm communicated the recall to customers who had received affected kits. Medline instructed customers to request stickers to over-label the affected convenience kits with a warning label, indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
Initiated
2025-07-08
Root cause
Unknown/Undetermined by firm
Distribution
US Nationwide distribution tin the states of CA, FL, IA, and TX.
Recall number
Z-2452-2025
Recall status
Open, Classified
CDRH record ID
215404
FDA classified
2025-08-27
Recall event number
97371

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.