Open original FDA record ↗Recall record
- FDA identifier
- Z-2453-2021
- Company in FDA record
- Accora Inc
- FDA record creation date
- Sep 13, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has identified a potential for the backrest to become detached from the chair.
Source facts
- Action
- On June 21, 2021, the firm sent affected customers Urgent Field Safety Notices via email. The notice identified the affected products and informed customers of the product issue. The corrective action will be to fit a modification kit to eliminate this fault condition and ensure the backrest cannot detach from the chair. All new production chairs will have this kit fitted as standard. Customers should make sure that all caregivers and users of the affected Configura Advance Chair devices are made aware of this Field Safety Notice. Furthermore, they should contact Accora to arrange a Field Safety Corrective Action kit to be installed on each product. This will be fitted free of charge. If you have questions or require assistance, please contact Accora: Accora Limited, Charter House, Barrington Road, Orwell, Cambridge SG8 5QP T: +44 (0)1223 206100 E: info@accora.care
- Initiated
- 2021-06-21
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide in the states of AL, FL, IL, MN, MS, NH, NJ, NY, NC, OH, PA, SC and the countries of Ireland, UK, and Australia.
- Recall number
- Z-2453-2021
- Recall status
- Open, Classified
- CDRH record ID
- 188670
- FDA record created
- 2021-09-13
- Recall event number
- 88441