Z-2455-2024: DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures anal

Recall record Z-2455-2024 for Beckman Coulter Mishima K.K.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2455-2024
Company in FDA record
Beckman Coulter Mishima K.K.
FDA classification date
Jul 31, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient order and a calibration order. When a patient sample and a non-barcoded calibrator tube is processed on the rack, the calibration result will be reported as a patient result. Issue 2: calibration with expired calibrator. They both may cause erroneous patient results, but the probability of serious adverse health consequences or death is unlikely.

Source facts

Action
During the week of June 20, 2024, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter via email/postal mail to customers informing them of two software related issues associated with use of the Clinical Chemistry Analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient order and a calibration order. When a patient sample and a non-barcoded calibrator tube is processed on the rack, the calibration result will be reported as a patient result. This may cause a false low result or a false high result to be reported to a physician. Issues 2: When a calibration order is manually placed (calibrator lot is not expired) and then the calibrator expires before the calibrator material is placed onto the analyzer for processing, the analyzer will process the expired calibration lot and generate patient results that may be inaccurate, potentially causing a false low result or a false high result to be reported to a physician. Customer are instructed to: For Issue 1 -Follow the IFU Note in Assigning Racks or Calibrator IDs to Calibrators section on pg. 8-7. -Alternatively, change calibrator Load Type to Bar Code Configuration and use barcoded containers for calibration. -Refer to IFU Assigning Racks or Calibrator IDs to Calibrators section on pg. 8-6, and in the software user interface Calibration section, under Calibrations Due Now, review the Load Type configuration, rack ID and cup position set up for the calibrators. Issue 2: -Verify that the calibrator has not expired prior to loading the calibrator material onto the analyzer for processing. -Cancel calibration orders after the calibrator lot material expires. -Repeat any test results generated from a calibration curve that used expired calibrator lot material. For questions/assistance-contact Customer Support Center at http://www.beckmancoulter.com
Initiated
2024-06-05
Root cause
Software design
Distribution
Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, and South Africa.
Recall number
Z-2455-2024
Recall status
Open, Classified
CDRH record ID
208405
FDA classified
2024-07-31
Recall event number
94820

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗