Skip to content

Public FDA record

Z-2456-2026: TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816

Recall record Z-2456-2026 for Smith & Nephew, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2456-2026
Company in FDA record
Smith & Nephew, Inc.
FDA classification date
Jun 16, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.

Source facts

Action
On 3/16/2026, recall notices were emailed to customers who were asked to do the following: 1) Locate and quarantine affected devices immediately. If you have further distributed the product to other organizations, please inform them at once of this Field Action and provide to them a copy of this letter. 3. Please complete the Customer Response form and email or fax it to your national Smith+Nephew agency/distributor. 4. Return quarantined product to your national Smith+Nephew agency/distributor. 5. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action. Firm contact information: + 1 901 396 2121, 1 800 821 5700
Initiated
2026-03-16
Root cause
Process control
Distribution
International distribution to the countries of BE, DE, IT.
Recall number
Z-2456-2026
Recall status
Open, Classified
CDRH record ID
219265
FDA classified
2026-06-16
Recall event number
98629

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗