Z-2457-2008: Cholestech LDX High Sensitivity C-Reactive Protein (hs-CRP) Test Cassette. The Cholestech LDX high sensi

Recall record Z-2457-2008 for Cholestech Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2457-2008
Company in FDA record
Cholestech Corp
FDA classification date
Sep 20, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Test results are high, outside of control material upper limit specification. Results could be as high as 15%.

Source facts

Action
Cholestech provided its three distributors a notification letter/response form with instructions to distribute the letter to their customers/end users dated August 9, 2006. The letter instructed customers to discontinue use of the lot of cassettes, return remaining cassettes to Cholestech through the Technical Service Representative and to complete and return the response form. The end users returned the signed letters directly to Cholestech so that effectiveness checks could be completed.
Initiated
2006-08-09
Root cause
Device Design
Terminated
2009-09-30
Distribution
Nationwide Distribution --- including states of Virginia, Texas and North Carolina, Ohio, Arkansas, Washington, Florida , California and Michigan.
Recall number
Z-2457-2008
Recall status
Terminated
CDRH record ID
72854
FDA classified
2008-09-20
Recall event number
49142

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗