Open original FDA record ↗Recall record
- FDA identifier
- Z-2457-2008
- Company in FDA record
- Cholestech Corp
- FDA classification date
- Sep 20, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Test results are high, outside of control material upper limit specification. Results could be as high as 15%.
Source facts
- Action
- Cholestech provided its three distributors a notification letter/response form with instructions to distribute the letter to their customers/end users dated August 9, 2006. The letter instructed customers to discontinue use of the lot of cassettes, return remaining cassettes to Cholestech through the Technical Service Representative and to complete and return the response form. The end users returned the signed letters directly to Cholestech so that effectiveness checks could be completed.
- Initiated
- 2006-08-09
- Root cause
- Device Design
- Terminated
- 2009-09-30
- Distribution
- Nationwide Distribution --- including states of Virginia, Texas and North Carolina, Ohio, Arkansas, Washington, Florida , California and Michigan.
- Recall number
- Z-2457-2008
- Recall status
- Terminated
- CDRH record ID
- 72854
- FDA classified
- 2008-09-20
- Recall event number
- 49142