Open original FDA record ↗Recall record
- FDA identifier
- Z-2457-2015
- Company in FDA record
- RAYSEARCH LABORATORIES AB
- FDA classification date
- Aug 25, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior. These contraction distances are, in error, pairwise reversed, i.e. right is interpreted as left, inferior is interpreted as superior, and posterior is interpreted as anterior.
Source facts
- Action
- On May 27, 2015 RaySearch Laboratories distributed Medical Device Correction notices to their customers via courier service. Within the Medical Device Correction notice, RaySearch Laboratories included a workaround to the problem for their customers. Customers should not use structure templates containing derived ROI definitions with non-uniform contractions. They should avoid using non-uniform contractions. In the case non-uniform contractions must be used, enter switched distances manually and review the result carefully before using it for treatment planning. Lastly, do not use humerus shielding in Automatic Breast Planning. Customers are also expected to complete and return the lower portion of the correction notice via fax to 888-501-7195. **CAPA** Correct the software problem in patch version 4.7.2 of RayStation 4.7, scheduled for market release June 2015.
- Initiated
- 2015-05-27
- Root cause
- Software design
- Terminated
- 2017-09-22
- Distribution
- Distributed in CO, IL, MI, NC, OH,TN, and WA.
- Recall number
- Z-2457-2015
- Recall status
- Terminated
- CDRH record ID
- 137704
- FDA classified
- 2015-08-25
- Recall event number
- 71414