Skip to content

Public FDA record

Z-2457-2024: Arctic Sun Temperature Management Systems, Name/REF: Arctic Sun STAT, NA/6000-00-00; Arctic Sun STAT Loa

Recall record Z-2457-2024 for Medivance Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2457-2024
Company in FDA record
Medivance Inc.
FDA classification date
Jul 31, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue.

Source facts

Action
On 6/27/24, correction notices were sent to customers and distributors informing them of the following: Advice for Clinical Users: 1) Systems contains the following two fixed alarms when the device is operating in patient control mode: - Alarm 10 will alert if the patient s temperature and/or water temperature is below 31 degrees Celsius and the water temperature is below 31 degrees Celsius. - Alarm 08 will alert if the patient s temperature and/or water temperature is above 39.5 degrees Celsius. 2) Customers should continue to use the customizable Patient Temperature Low (Alarm 11) and Patient Temperature High (Alarm 09) alarms on their devices. Both alarms are available when the device is operating in patient control mode. Customers should set these alarms to the lowest and highest temperatures they feel are acceptable for the specific patient being treated. Default settings are 30 degrees Celsius and 44 degrees Celsius and alert independent of water temperature. For Arctic Sun STAT, refer to the Operator Manual for Alarm 9 on page 39 and Alarm 11 on page 40. For Arctic Sun 5000, refer to the Service Manual for Alarms 9 and 11 on page 9. 3) As instructed in the Operating Manual, always empty Arctic Gel Pads when therapy is ended, even if the stop in therapy is only temporary. Failure to do so may result in overfill of the water reservoir and this can affect temperature management. -Complete and return the response form via email to BDRC40@bd.com -Share this notice with any users of the product within your facilities or with any interfacility users where product was transferred, to ensure they are also aware of this Medical Device Recall. -Firm is developing a software update and will provide an additional communication with the status of the software update release by February 28, 2025. If you have any questions, contact the firm's Medical Information Services that are available 24/7 at 1-866-840-9776.
Initiated
2024-06-27
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution including in the states of IN, OK, MO, FL, CT, KS, TX, NJ, CA, OR, HI, IL, WI, VA, KY, NH, CO, NV, LA, NC, ID, UT, GA, SC, TN, MD, PA, IA, MA, OH, NY, DE, MI, NM, WV, AZ, DC, VT, MN, SD, AL, AR, WA, NE, ND, MT, AK, MS, ME, WY, RI, PR, OK, OR and the countries of CA, DE, AT, BE, FI, TR, NO, DK, JP, ES, AU, FR, GB, IT, SE, PT, AE, SA, ZA, RU, KR, CY, TH, GR, TW, MY, OM, CN, SG, GI, NL, IL, CZ, RO, CH, RE, PL, PE, SK, AR, IN, CL, MX, QA, BR, SI, HU, JO, LU, KW, PH, NZ, HK, VN, IR, BN, UA, ID, CO, KZ, HR, IS, GU, GE, BH, EE, AM, PA, KR, CL, TH, MY, JP, AU, BD, BE, TW
Recall number
Z-2457-2024
Recall status
Open, Classified
CDRH record ID
208570
FDA classified
2024-07-31
Recall event number
94881

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.