Open original FDA record ↗Recall record
- FDA identifier
- Z-2458-2008
- Company in FDA record
- Conmed Linvatec Endoscopy Division
- FDA classification date
- Sep 21, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated because ConMed Linvatec has determined there is a possibility the wheel caster(s) may become loose which may result in the wheel caster(s) falling out of the base of the VP8500 Video Cart. The potential for the wheel caster or casters to fall out of the base of the cart may result in a potential tip hazard.
Source facts
- Action
- The recall was initiated by the firm sending the recall information packet via UPS overnight delivery to the Distribution and Sales Force in May 2008. The packet included: 1. Recall Notification and reply form for each consignee. 2. Instructions on the procedure to conduct the recall of the VP8500. 3. A list of affected consignees in their territory. The Recall Notification informed the customers of the potential for the wheel caster(s) to fall out of the base of the cart which may result in a potential tip hazard. The customers were instructed to discontinue the use of the VP8500 Video Cart immediately if the wheel caster(s) show any signs of being loose and to complete the recall reply form. Upon receipt of notification, distributors are supposed to : 1. Visit the affected accounts and provide them with a copy of the Recall Notification to inform them of the recall and the Recall Acknowledgment Reply Form which is to be mailed or faxed back to the firm. 2. Schedule the delivery and installation of refurbished carts to replace the recalled cart(s) by contacting Jill Cover, ConMed Linvatec Endoscopy Division at (888) 292-0100, ext. 8157 or e-mail jcover@livatec.com to arrange for replacement. 3. Move medical devices, etc. from the recalled cart(s) and set up the refurbished cart(s) with the required equipment at the account's location. (ProMEDICA will arrange for shipping of the recalled cart(s) back to ProMEDICA.)
- Initiated
- 2007-10-31
- Root cause
- Device Design
- Terminated
- 2008-09-22
- Distribution
- Worldwide Distribution including states of: AK, AZ, CA, FL, GA, IL, IN, KS, MD, NC, NY, OH, and PA and country of Canada.
- Recall number
- Z-2458-2008
- Recall status
- Terminated
- CDRH record ID
- 70983
- FDA classified
- 2008-09-21
- Recall event number
- 48195