Open original FDA record ↗Recall record
- FDA identifier
- Z-2458-2026
- Company in FDA record
- AMD Medicom Inc.
- FDA classification date
- Jun 16, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
Source facts
- Action
- On 5/8/2026, recall notices were emailed to customers who were asked to do the following: 1) Discontinue use, quarantine and destroy affected devices. 2) Notify any customers or facilities to whom you have distributed this product. 3) Complete and return the response form via email to firm's customer service at Claims@medicom.ca On 6/5/2026, recall notices with all affected lots were emailed to customers.
- Initiated
- 2026-05-08
- Root cause
- Material/Component Contamination
- Distribution
- US: MT, NC, VA, CA, TX, OH, TN, IL, RI
- Recall number
- Z-2458-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220498
- FDA classified
- 2026-06-16
- Recall event number
- 99006