Reason stated in the source
NicoletOne 5.30.2 Software when used with CSeries amplifiers in Fixed Room LTM Configurations with patient room kits. Cardinal Health NeuroCare has confirmed that the 3rd party room relay connected to the A2 wall plate does not always activate if the patient event button connected to the CSeries Amplifier is pressed twice within one second.
Source facts
- Action
- Consignees were sent a "Cardinal Health Urgent: Medical Device Correction" letter dated June 23, 2008. The letter identified the product and problem. It also provided instructions (Field Correction) to install the software patch and requested return of the verification form. Contact Cardinal Health NeuroCare Division at 1-608-271-3333 for assistance.
- Initiated
- 2008-06-13
- Root cause
- Software design
- Terminated
- 2011-12-11
- Distribution
- US: AL, AR, CA, CO, FL, GA, IL IN, KY, LA, MO MS, NC, NJ, NY, OH, PA, TN, TX, WA, and WI. OUS: CH, CN, DK, GB, HK, IN, NO, RU, and UK.