Skip to content

Public FDA record

Z-2459-2008: NicoletOne 5.30.2 Software when used with CSeries amplifiers in Fixed Room LTM Configurations with patie

Recall record Z-2459-2008 for Cardinal Health NeuroCare Division. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2459-2008
Company in FDA record
Cardinal Health NeuroCare Division
FDA classification date
Sep 23, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

NicoletOne 5.30.2 Software when used with CSeries amplifiers in Fixed Room LTM Configurations with patient room kits. Cardinal Health NeuroCare has confirmed that the 3rd party room relay connected to the A2 wall plate does not always activate if the patient event button connected to the CSeries Amplifier is pressed twice within one second.

Source facts

Action
Consignees were sent a "Cardinal Health Urgent: Medical Device Correction" letter dated June 23, 2008. The letter identified the product and problem. It also provided instructions (Field Correction) to install the software patch and requested return of the verification form. Contact Cardinal Health NeuroCare Division at 1-608-271-3333 for assistance.
Initiated
2008-06-13
Root cause
Software design
Terminated
2011-12-11
Distribution
US: AL, AR, CA, CO, FL, GA, IL IN, KY, LA, MO MS, NC, NJ, NY, OH, PA, TN, TX, WA, and WI. OUS: CH, CN, DK, GB, HK, IN, NO, RU, and UK.
Recall number
Z-2459-2008
Recall status
Terminated
CDRH record ID
73198
FDA classified
2008-09-23
Recall event number
49350

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗