Skip to content

Public FDA record

Z-2459-2026: Ritmed Neurological Sponges, 0.75x0.75in, Sterile, REF: 85995C

Recall record Z-2459-2026 for AMD Medicom Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2459-2026
Company in FDA record
AMD Medicom Inc.
FDA classification date
Jun 16, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.

Source facts

Action
On 5/8/2026, recall notices were emailed to customers who were asked to do the following: 1) Discontinue use, quarantine and destroy affected devices. 2) Notify any customers or facilities to whom you have distributed this product. 3) Complete and return the response form via email to firm's customer service at Claims@medicom.ca On 6/5/2026, recall notices with all affected lots were emailed to customers.
Initiated
2026-05-08
Root cause
Material/Component Contamination
Distribution
US: MT, NC, VA, CA, TX, OH, TN, IL, RI
Recall number
Z-2459-2026
Recall status
Open, Classified
CDRH record ID
220510
FDA classified
2026-06-16
Recall event number
99006

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗