Z-2460-2008: Cholestech GDX A1C Test Cartridge. The Cholestech GDX A1C Test is an in-vitro diagnostic test that measu

Recall record Z-2460-2008 for Cholestech Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2460-2008
Company in FDA record
Cholestech Corp
FDA classification date
Sep 23, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Cholestech Corp. was notified by the foreign manufacturer of the A1C test cartridge that stability data indicated that the product would not meet performance claims through the end of its shelf life.

Source facts

Action
A customer notification letter was sent to Cholestech consignees who had received product from this lot. The letter dated 8/3/2006 instructed customers to discontinue use of the lot of product, destroy product according to in-house standard procedures and contact Cholestech Technical Services for replacement. It also instructed to provide a copy of the letter to other users if the lot of product was further distributed. A response form was included with instructions to complete and return. Contact Cholestech Corp at 1-510-781-3505 for assistance.
Initiated
2006-08-03
Root cause
Nonconforming Material/Component
Terminated
2009-10-07
Distribution
U.S. state of Colorado. OUS: India, Slovenia, Germany, Tunisia, Taiwan, Canada, Philippines, South Africa and Mexico
Recall number
Z-2460-2008
Recall status
Terminated
CDRH record ID
73203
FDA classified
2008-09-23
Recall event number
49359

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗