Open original FDA record ↗Recall record
- FDA identifier
- Z-2460-2012
- Company in FDA record
- Alere San Diego, Inc.
- FDA classification date
- Sep 26, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Alere San Diego is recalling Alere Triage TOX Drug Screen because there is a possibility that after installing the code chip provided with the affected lots: AMP, BAR, and PCP, the threshold ranges may not be consistently set correctly in all meters.
Source facts
- Action
- Alere sent a 'NOTICE OF CORRECTION" letter dated July 12, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers with questions were instructed to contact Technical Services at (877) 308-8287. Customers were instructed to complete the Customer/Distributor Verification Form Notice of Correction even if they do not have any involved product and fax back to Technical Services at (858) 805-8457 or scan and email to responses.ts@alere.com.
- Initiated
- 2012-07-12
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-02-01
- Distribution
- Nationwide Distribution.
- Recall number
- Z-2460-2012
- Recall status
- Terminated
- CDRH record ID
- 112528
- FDA classified
- 2012-09-26
- Recall event number
- 63091