Z-2460-2012: Alere Triage TOX Drug Screen 9 Panel, P/N: 94402. The Alere Triage TOX Drug Screen is a fluorescence i

Recall record Z-2460-2012 for Alere San Diego, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2460-2012
Company in FDA record
Alere San Diego, Inc.
FDA classification date
Sep 26, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Alere San Diego is recalling Alere Triage TOX Drug Screen because there is a possibility that after installing the code chip provided with the affected lots: AMP, BAR, and PCP, the threshold ranges may not be consistently set correctly in all meters.

Source facts

Action
Alere sent a 'NOTICE OF CORRECTION" letter dated July 12, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers with questions were instructed to contact Technical Services at (877) 308-8287. Customers were instructed to complete the Customer/Distributor Verification Form Notice of Correction even if they do not have any involved product and fax back to Technical Services at (858) 805-8457 or scan and email to responses.ts@alere.com.
Initiated
2012-07-12
Root cause
Nonconforming Material/Component
Terminated
2013-02-01
Distribution
Nationwide Distribution.
Recall number
Z-2460-2012
Recall status
Terminated
CDRH record ID
112528
FDA classified
2012-09-26
Recall event number
63091

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗