Z-2461-2025: IPC Powerease System REF 2300000 UDI-DI: 00613994448705 00643169406834. The IPC POWEREASE System is indi

Recall record Z-2461-2025 for Medtronic Xomed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2461-2025
Company in FDA record
Medtronic Xomed, Inc.
FDA classification date
Aug 27, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to out of the box wobble of the driver.

Source facts

Action
On 07/17/2025, the firm initiated distribution of "URGENT: MEDICAL DEVICE RECALL" Letters via email and/or mail to customers informing them of reported out-of-box failures due to device wobble which is due to misalignment between the quick connect and shaft main driver components of the IPC Powerease System Model 2300000. Customers are instructed to: . Identify, segregate, quarantine and stop use of any affected products within your inventory. The list of affected device serial numbers is included in Attachment A: Product Serial Number List. . Your Medtronic field representative will coordinate the replacement of your affected product. . Return affected product in your inventory to Medtronic. Please see Customer Confirmation Form for instructions on returning impacted devices. . Please complete and return the customer confirmation form enclosed with this letter acknowledging receipt of this information via email to neuro.quality@medtronic.com even if you no longer have possession, custody, or control over the affected product. Note: Instructions on returning the acknowledgement form to Medtronic is located on the form itself. . Please share this communication within your organization, with other organizations where impacted devices have been transferred, and any other associated organizations that may be impacted by this action. Maintain a copy of this letter for your records. For questions or assistance please contact Medtronic Sales Representative and/or Technical Services at 1-888-826-5603.
Initiated
2025-07-17
Root cause
Process change control
Distribution
U.S.: AZ, CA, CO, CT, FL, GA, IA, IL, IN, LA, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OR, SC, TN, TX, O.U.S.: Canada, China, Croatia, France, Germany, India, Italy, Japan, Jordan, Kuwait, Norway, Romania, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom
Recall number
Z-2461-2025
Recall status
Open, Classified
CDRH record ID
215013
FDA classified
2025-08-27
Recall event number
97266

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗