Varian has identified an anomaly with the Eclipse Treatment Planning where an outdated assigned Hounsfield Unit [HU] value can be used for dose calculation when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server. Treatment of the patient using these values can therefore lead to under or over-dose.
Source facts
Action
VARIAN Medical Systems sent an Urgent Medical Device Correction Urgent Field Safety Notice on August 17, 2012, to all affected customers with a description of the problem and user corrective action steps. Customers were instructed to ensure that the workstation clock be synchronized with the DatabaseSystem server clock. Customers were also instructed to contact Varian service for assistance with time synchronization if required. Customers were to complete the attached Proof of Notification Form once they have read the document and return it to Varian Medical Systems.
For questions customers should call the following numbers:
USA and Canada 1-888-827-4265.
Europe +41 41 749 8844.
For questions regarding this recall call 650-424-5731.
Initiated
2012-08-17
Root cause
Software design
Terminated
2014-08-06
Distribution
Worldwide Distribution - USA (nationwide) and Internationally.
.
.
Recall number
Z-2462-2012
Recall status
Terminated
CDRH record ID
112531
FDA classified
2012-09-26
Recall event number
63093
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.