Z-2463-2026: Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 an

Recall record Z-2463-2026 for Hamilton Medical AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2463-2026
Company in FDA record
Hamilton Medical AG
FDA classification date
Jun 30, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement

Source facts

Action
On 5/29/2026, correction notices were emailed to customers who were asked to do the following: 1) Inform all potential users of affected devices within your facility about this issue. 2) Inform all affected end users and/or sub-distributors and provide them with this notice. 5) Post this notice to ensure all personnel are informed. 6) Complete and return Facility Reply Form via email to complaints@hamiltonmedical.com Customers with questions can contact the firm at the email above. Initially customers were informed to continue use and perform actions to mitigate risk, but later the updated notice called for affected device removal. On 7/8/2026, updated recall notices were emailed and mailed to healthcare facilities, physicians and end users who were informed of the following: 1) Discontinue use and quarantine affected devices. 2) Do not follow previous instructions for product modification. 3) Inform all affected end users and/or sub-distributors and provide them with this notice. 4) Contact firm if you have affected product in inventory. - Discard any affected product(s) and confirm destruction. Or - Contact Hamilton Medical for a return material authorization number and return label. -Contact your sales representative to acquire replacement product. 5) Complete and return the Facility Reply Form.
Initiated
2026-05-29
Root cause
Component design/selection
Distribution
US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.
Recall number
Z-2463-2026
Recall status
Open, Classified
CDRH record ID
220606
FDA classified
2026-06-30
Recall event number
99028

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗