Open original FDA record ↗Recall record
- FDA identifier
- Z-2464-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Sep 26, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the head-foot direction, the coil can heat up in the area of the electronic housing (white plastic lids).
Source facts
- Action
- Siemens Healthcare sent a Customer Safety Advisory Notice dated September 5, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter included updated instruction to all customers. For questions regarding this recall call 610-219-6300.
- Initiated
- 2012-09-12
- Root cause
- Device Design
- Terminated
- 2014-01-14
- Distribution
- Nationwide Distribution - USA including Washington, DC and AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI and Puerto Rico.
- Recall number
- Z-2464-2012
- Recall status
- Terminated
- CDRH record ID
- 112954
- FDA classified
- 2012-09-26
- Recall event number
- 63197