Open original FDA record ↗Recall record
- FDA identifier
- Z-2464-2023
- Company in FDA record
- RAYSEARCH LABORATORIES AB
- FDA classification date
- Aug 25, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstances
Source facts
- Action
- RaySearch issued Field Safety Notice, Medical Device Correction #119464 Letter via E-mail, on July 17, 2023. Letter states reason for recall, health risk and action to take: Manually verify the patient s medication list with respect to included substances when adding an new allergy to a medication substance. " Make sure to update all medications when activating a new version of the drug ingredients value set in the RayCare admin interface. " Educate staff and all users about the absence of a warning when adding a patient allergy. " Inspect your product and identify all installed units with the above software version number(s). " Confirm that you have read and understood this notice by replying to the notification email. SOLUTION This issue is resolved in the RayCare 2023B, market released in July 2023 (subject to market clearance in some markets) and future versions. If customers wish to continue using versions of RayCare affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. For regulatory information, please contact quality@raysearchlabs.com.
- Initiated
- 2023-07-17
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the state of TN.
- Recall number
- Z-2464-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202525
- FDA classified
- 2023-08-25
- Recall event number
- 92777