The hook end effecter from a monopolar device may become dislodged and fall into the patient when in use during laproscopic surgery. Retrieval from the patient may involve extended anesthesia time.
Source facts
Action
TransEnterix issued a "Field Action" letter dated June 22, 2010 to consignees, identifying the affected product and actions to be taken by the customer. Customers were instructed to return the product to TransEnterix, Inc. using the address label and return information provided. The firm will replace the product at no charge. TransEnterix Customer Care can be contacted at 1-888-879-4111.
Initiated
2010-06-22
Root cause
Other
Terminated
2011-01-31
Distribution
Nationwide Distribution in the states of AZ, FL, GA, NC, NY, OH, OR, SC and TX.
Recall number
Z-2465-2010
Recall status
Terminated
CDRH record ID
92858
FDA classified
2010-09-22
Recall event number
56179