Open original FDA record ↗Recall record
- FDA identifier
- Z-2465-2012
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Sep 27, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The Wireless Portable Detector (WPD) is not fixed securely inside the Mobile Detector Holder and may fall down
Source facts
- Action
- Philips Healthcare sent a Urgent Field Safety Notice dated August 13, 2012, to all affected customers. Notifications began the week of September 6, 2012, for US and Worldwide Customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips field service engineer will Secure the 4 countersunk screws, which are used to attach the lock mechanism, with thread locker (loctite). Further information or support concerning this issue, contact your local Philips representative. 1-800-722-9377.
- Initiated
- 2012-09-06
- Root cause
- Device Design
- Terminated
- 2017-01-17
- Distribution
- Worldwide Distribution--USA (nationwide) and the countries of Australia, Austria, Canada, Chile, Finland, France, Germany, Hungary. Italy, Japan, Latvia, Netherlands, New Zealand. Oman, South Korea, Switzerland and United Kingdom.
- Recall number
- Z-2465-2012
- Recall status
- Terminated
- CDRH record ID
- 112650
- FDA classified
- 2012-09-27
- Recall event number
- 63119