Skip to content

Public FDA record

Z-2467-2023: EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V

Recall record Z-2467-2023 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2467-2023
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Aug 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Recent reports of patient infection.

Source facts

Action
Consignees were sent an Urgent Medical Corrective Action notification dated 7/27/2023. This action is being initiated due to the firm becoming aware of and FDA inquiring about recent reports of infections and positive cultures. Olympus is reminding users to closely follow reprocessing instructions, including periodic inspections by Olympus Service, and informing users about an updated manual intended to improve comprehension of existing reprocessing steps. Additional enhancements to the manual include the use of reprocessing timelines and simplified text/visualization techniques. The manual is available on the customer web portal and paper copies, along with wall charts, are available to be mailed to your facility. The notification also emphasises the required inspection of the forceps elevator mechanism, to be performed once a year or every 100 reprocessing cycles, and encourages consignees to consider incorporating voluntary routine microbiologic surveillance of endoscops as part of their quality improvement/assurance efforts. Consignees are asked to complete and return the provided response form to Sedgwick via email at ollympus7521@sedgwick.com or by fax to 888-208-4588. Questions can be directed by phone to 866-912-9552. Customers can also contact Ashley Mitch at Olympus by phone at 484-553-1029 or by email at Ashley.Mitch@olympus.com with any additional concerns.
Initiated
2023-07-27
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-2467-2023
Recall status
Open, Classified
CDRH record ID
202210
FDA classified
2023-08-28
Recall event number
92703

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗