Skip to content

Public FDA record

Z-2468-2020: ABL80-FLEX CO-OX 393-841 Box Label, REF 393-841, IVD, CE, UDI: 05700693938417 - Product Usage: It is int

Recall record Z-2468-2020 for Radiometer Medical ApS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2468-2020
Company in FDA record
Radiometer Medical ApS
FDA record creation date
Jun 29, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.

Source facts

Action
The firm will send consignees Urgent Medical Device Recall Letter dated 5/11/2020 via email and FedEx 2nd day delivery with proof of delivery on 5/11/2020 informing them that Radiometer has become aware that the internal barcode reader, included in every ABL80 FLEX Series analyzers, may potentially misinterpret the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. The issue identified only relates to barcode types not using a check digit. A check digit enables the barcode reader to validate the data read, and hence, to capture if the barcode has been misinterpreted and consequently to reject such data. For barcodes without a check digit, the following factors may add to the risk of misinterpretation: " Poor paper quality " Poor printer quality " Improper handling of the barcode (e.g. the barcode is folded or exposed to liquid) Radiometer uses only barcode types with a check digit when producing barcode labels used for e.g. sampler identification. They request customers to take the following actions: Check if their institution is using barcode types without a check digit, including e.g. patient ID or accession number, to be read on the ABL80. 1. If you use barcode types without a check digit, Radiometer recommends to either: o Enable the check digit on the barcode type currently used, or o Change the type of barcode to one that includes a check digit, as this enables the barcode reader to validate the data read, and hence, to recognize if the barcode has been misinterpreted and consequently to reject such data.2. If you use barcode types with a check digit, the barcode reader already validates the data read and is able to identify if the barcode has been misinterpreted and consequently rejects such data. Hence, no short-term action is required. 3. If you do not use locally generated barcodes to be read by the ABL80, no short-term action is required.
Initiated
2020-05-11
Root cause
Under Investigation by firm
Terminated
2023-02-24
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AK, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CA.
Recall number
Z-2468-2020
Recall status
Terminated
CDRH record ID
181429
FDA record created
2020-06-29
Recall event number
85634

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗