Open original FDA record ↗Recall record
- FDA identifier
- Z-2470-2012
- Company in FDA record
- Alcon Research, LTD.
- FDA classification date
- Sep 27, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
This medical device correction is related to Alcon products that present a 23G non-valved trocar cannula and infusion cannula together because of an issue regarding the loose fit between the 23G non-valved trocar cannula and infusion cannula that could potentially cause a detachment between the non-valved trocar cannula and infusion cannula during surgery.
Source facts
- Action
- Alcon sent an Urgent a Voluntary Medical Device Correction letter on May 9, 2012. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed NOT to use the affected product, instead customers should use a 1-Count Valved Trocar Cannula and to complete and immediately return the Medical Device Correction Response Form to Alcon via fax at: (817) 916-9087 or email to: ReplacementCannulas@alconlabs.com. For questions customers should call 1-800-862-5266.
- Initiated
- 2012-05-29
- Root cause
- Device Design
- Terminated
- 2013-02-12
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, United Kingdom, Greece, Singapore, Germany, Switzerland, Kuwait, United Arabian Emirates, and Austria.
- Recall number
- Z-2470-2012
- Recall status
- Terminated
- CDRH record ID
- 110462
- FDA classified
- 2012-09-27
- Recall event number
- 62419