Open original FDA record ↗Recall record
- FDA identifier
- Z-2470-2018
- Company in FDA record
- Ion Beam Applications S.A.
- FDA record creation date
- Jul 17, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Gantry rotation or proton irradiation can be allowed even if apertures or accessories are not correctly positioned in snout 300x400 or snout XL. This can lead to risk of crushing for patient and/or user, and risk of mistreatment for the patient.
Source facts
- Action
- Recall was initiated on 6/27/18. Firm plans to provide the Urgent Medical Device Correction notifications to IBA site managers present on customer sites. The site managers will deliver the notification to customers by hand or email. The recall letter alerts users to the potential issue and instructs users to verify that apertures and accessories are fully inserted with both sides in the guide. The firm plans to deploy a solution on or before May 31, 2019, that will prevent occurrences of the issue.
- Initiated
- 2018-06-27
- Root cause
- Device Design
- Terminated
- 2020-05-20
- Distribution
- US Nationwide Distribution in the states of FL, VA, PA, OK, NJ, WA, and TX.
- Recall number
- Z-2470-2018
- Recall status
- Terminated
- CDRH record ID
- 165327
- FDA record created
- 2018-07-17
- Recall event number
- 80384