Open original FDA record ↗Recall record
- FDA identifier
- Z-2470-2020
- Company in FDA record
- Siemens Healthcare Diagnostics Inc.
- FDA record creation date
- Jun 29, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Flexible tubing connected to the liquid waste bottle can crack during routine customer maintenance potentially causing liquid waste can leak onto the floor, creating a slip and fall hazard
Source facts
- Action
- Siemens issued an Urgent Medical Device Correction (UMDC IMI20-01.A.US) to US customers via FedEx on May 15, 2020. Urgent Field Safety Notice (UFSN IMI20-01.A.OUS) was electronically issued to all OUS Siemens sites with affected customers, for distribution, as well as regional reporting to their respective local competent authority on May 15, 2020. The field action letters explain the issue, health risk and and inform customers of actions to take: Siemens Customer Service Engineers shall be visiting customer sites to replace Waste Bottle Flex Tubing Assemblies. Customers are requested to return the Effectiveness Check Form, in the UMDC/UFSN via fax or email.
- Initiated
- 2020-05-15
- Root cause
- Nonconforming Material/Component
- Terminated
- 2022-08-12
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of AE, AO, AR, AT, AU, AZ, BA, BD, BE, BG, BH, BO, BR, BS, BY, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GP, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KR, KW, KZ, LB, LT, LU, LV, LY, ME, MK, MM, MQ, MT, MU, MX, MY, NL, NO, NZ, PE, PH, PK, PL, PT, RO, RS, RU, SA, SE, SG, SI, SK, SY, TH, TN, TR, TT, TW, UA, UY, VE VN, ZA.
- Recall number
- Z-2470-2020
- Recall status
- Terminated
- CDRH record ID
- 181653
- FDA record created
- 2020-06-29
- Recall event number
- 85731