Z-2470-2023: GS70H Salus Surgical Table, Product Code 18-071-70

Recall record Z-2470-2023 for Skytron, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2470-2023
Company in FDA record
Skytron, LLC
FDA classification date
Aug 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

Source facts

Action
An URGENT: Medical Device Correction dated 7/25/23 was sent to customers. Actions to be taken by the user: 1. Please promptly forward this notification if the device has been transferred to any other entity. 2. Please review appendices of affected serial and UDI numbers and identify devices within your facility. 3. Please do not use the table control in Bluetooth mode until the correction has been installed. 4. Please contact quality at quality@skytron.com with any questions. Skytron authorized service technicians will contact affected facilities to schedule this service. A location with a medical outlet and a power cord for the operating table available will be required to complete the corrective action. Skytron estimates that this correction activity will be completed by December 31, 2023. Please do not use the table control in Bluetooth mode until the correction has been installed.
Initiated
2023-08-01
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, CA, CT, FL, GA, IA, IL, IN, MA, MI, MO, PA, TN, TX, UT, WA, WI and the countries of Australia, New Zealand.
Recall number
Z-2470-2023
Recall status
Open, Classified
CDRH record ID
200148
FDA classified
2023-08-28
Recall event number
92290

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗