Open original FDA record ↗Recall record
- FDA identifier
- Z-2471-2020
- Company in FDA record
- Fujifilm Medical Systems U.S.A., Inc.
- FDA record creation date
- Jun 30, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Lateromedial (LM) and Lateromedial Oblique (LMO) Orientation Markers may be displayed incorrectly on the 3D TOMO slider bar. In addition, the slice location of the image and/or the slice direction could be incorrect as well, and result in misdiagnosis
Source facts
- Action
- Fujifilm issued URGENT: Medical Device Correction and Removal notification via FedEx standard overnight mail on May 22, 2020, stating reason for recall, health risk and action to take: recommending that you do not perform Tomosynthesis on these views until the correction is applied. FUJIFILM is providing hotfixes for Synapse PACS Versions 5.5, 5.7.0, and 5.7.1 to resolve these issues. Please contact Customer Service at 1-888-FUJIMED (1-888-385-4633). Cmplete and return the Field Action Verification Form Providing the response with the information requested is essential for ensuring appropriate action is taken. Contact field action coordinator, Jeffrey Wan, via phone at 201-675-8947 or email at jeffrey.wan@fujifilm.com, if you have any further questions regarding this field action.
- Initiated
- 2020-05-22
- Root cause
- Software design
- Distribution
- Nationwide
- Recall number
- Z-2471-2020
- Recall status
- Open, Classified
- CDRH record ID
- 181786
- FDA record created
- 2020-06-30
- Recall event number
- 85793