Skip to content

Public FDA record

Z-2471-2021: ACUSON Juniper Diagnostic Ultrasound System

Recall record Z-2471-2021 for Siemens Medical Solutions USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2471-2021
Company in FDA record
Siemens Medical Solutions USA, Inc.
FDA record creation date
Sep 14, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to intermittent failures of the power supply in the ultrasound system which renders it inoperable.

Source facts

Action
On 07/12/2021, the firm sent a "MEDICAL DEVICE SAFETY CORRECTION" Notification via email to customers informing them that there are intermittent failures of the power supply in the ultrasound system which may render the system inoperable. When an ultrasound system experiences this failure, cycling the power does not recover system functionality. Potential health risks include, lack of ultrasound guidance during interventional procedures (e.g., biopsy, guidance during surgery, patient under sedation) that rely on real-time ultrasound imaging to guide or complete the procedure. Customer will be contacted by a Customer Service Engineer to schedule a visit to their facility to update the system. Customers are asked to ensure that all users are aware of this issue and correction within their organization and others who may need to be inform of this situation. For further questions, contact Siemens Healthineers USD Complaints at usd-complaint.team@siemens-healthineers.com
Initiated
2021-07-12
Root cause
Software change control
Terminated
2024-10-16
Distribution
Worldwide distribution - U.S. Nationwide distribution in the Puerto Rico and the countries of Bosnia Herzegovina, Canada, Colombia, Germany, Greece, Italy, North Macedonia, Poland, and Slovenia.
Recall number
Z-2471-2021
Recall status
Terminated
CDRH record ID
188585
FDA record created
2021-09-14
Recall event number
88405

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗