Open original FDA record ↗Recall record
- FDA identifier
- Z-2471-2021
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA record creation date
- Sep 14, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to intermittent failures of the power supply in the ultrasound system which renders it inoperable.
Source facts
- Action
- On 07/12/2021, the firm sent a "MEDICAL DEVICE SAFETY CORRECTION" Notification via email to customers informing them that there are intermittent failures of the power supply in the ultrasound system which may render the system inoperable. When an ultrasound system experiences this failure, cycling the power does not recover system functionality. Potential health risks include, lack of ultrasound guidance during interventional procedures (e.g., biopsy, guidance during surgery, patient under sedation) that rely on real-time ultrasound imaging to guide or complete the procedure. Customer will be contacted by a Customer Service Engineer to schedule a visit to their facility to update the system. Customers are asked to ensure that all users are aware of this issue and correction within their organization and others who may need to be inform of this situation. For further questions, contact Siemens Healthineers USD Complaints at usd-complaint.team@siemens-healthineers.com
- Initiated
- 2021-07-12
- Root cause
- Software change control
- Terminated
- 2024-10-16
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the Puerto Rico and the countries of Bosnia Herzegovina, Canada, Colombia, Germany, Greece, Italy, North Macedonia, Poland, and Slovenia.
- Recall number
- Z-2471-2021
- Recall status
- Terminated
- CDRH record ID
- 188585
- FDA record created
- 2021-09-14
- Recall event number
- 88405