Skip to content

Public FDA record

Z-2473-2020: Elekta Unity, Image-Guided Radiation Therapy System

Recall record Z-2473-2020 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2473-2020
Company in FDA record
Elekta, Inc.
FDA record creation date
Jun 30, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There have been reports of excessive heating in the MRL Gradient Connection Module.

Source facts

Action
Initially, the firm issued an email on 12/18/2019 to their field engineers which said they had attached a "Knowledge Article" which provided instructions on how to inspect the Philips Marlin 1.5T system used in the Elekta Unity MR Linac. On 5/11/2020, the firm issued an Urgent Important Field Safety Notification dated May, 2020, via email to consignees which explained the problem, clinical impact, and said that initial corrective actions have been performed on all impacted sites to ensure the gradient coil connection assembly is correct. It goes onto say that extensive testing of the incorrect connection material has been completed to ensure that continued use of the Elekta Unity system is acceptable, however, it remained that further actions can be taken to reduce the risk.
Initiated
2019-12-18
Root cause
Process control
Terminated
2021-06-15
Distribution
Distributed in US - PA, NY, and IA. Distributed ROW - Australia, Canada, China, Switzerland, Denmark, Hong Kong, Italy, Netherlands, Sweden, Thailand, Turkey, UK. There was no government/military distribution.
Recall number
Z-2473-2020
Recall status
Terminated
CDRH record ID
181355
FDA record created
2020-06-30
Recall event number
85620

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗