Z-2474-2020: Flow Coupler Monitor, GEM, Global Excellence Microsurgery, Rx Only

Recall record Z-2474-2020 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2474-2020
Company in FDA record
Baxter Healthcare Corporation
FDA record creation date
Jun 30, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A firmware issue may cause the GEM1020M-2 Flow Coupler Monitor to intermittently shut down when a WiFi connection is not established.

Source facts

Action
Synovis Micro Companies Alliance (MCA), Inc., a subsidiary of Baxter International Inc., sent a notification to healthcare providers dated, June 2, 2020 titled, "Urgent Medical Device Recall". The notification was sent to the user level for the GEM1020M-2 Flow Coupler Monitor due to a firmware issue which may cause the monitor to intermittently shut down when a WiFi connection is not established. If a shutdown occurs, the monitor can be manually powered back on and will resume normal operation. Synovis is requesting the return of the monitors to perform a firmware upgrade to resolve the issue. Customers should contact Synovis to arrange for product return for the firmware upgrade and replacement of the product. Customers should complete the Baxter customer reply form and return it to Baxter by scanning and e-mailing it to fca@baxter.com, even if customers do not have any inventory. Customers should return the response form promptly. If customers have distributed the product to other facilities or departments within the institution, a copy of the communication should be forwarded to them. Questions regarding the communication, contact Synovis at 800-510-3318 or 205-941-0111, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday.
Initiated
2020-06-02
Root cause
Device Design
Distribution
U.S. distribution only: VA, GA, MD, PA, DC, WI, LA, MN, FL, IN, MI, AZ, TX, MS, MA, NY, SC, WI, WA, MI, CA.
Recall number
Z-2474-2020
Recall status
Open, Classified
CDRH record ID
181768
FDA record created
2020-06-30
Recall event number
85782

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗