Z-2474-2021: The Allura Xper series is intended for use on human patients to perform: "Vascular, cardiovascular and ne

Recall record Z-2474-2021 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2474-2021
Company in FDA record
Philips North America Llc
FDA record creation date
Sep 14, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.

Source facts

Action
Philips notified accounts by Urgent Medical Device Recall letter via certified mailing United States Postal Service (USPS) on 8/13/21. Letter states reason for recall, health risk and action to take: Please place this Medical Device Recall letter with the documentation of the system until this correction has been implemented in your system. Additionally, please return the attached reply form to Phhilps to confirm that the users of the system have reviewed and understood this Medical Device Recall. Philips will correct the affected systems by replacing the ADUs. A Philips service representative will contact you to schedule the correct ion of your system. Please be assured that maintaining a high level of safety and quality is our highest priority. lf you need any further information or support concerning thls issue, please contact your local Philips representative: 1-800-722-9377 (reference FCO72200483
Initiated
2021-08-13
Root cause
Device Design
Terminated
2024-03-11
Distribution
US Nationwide distribution in the states of ID, MA, MI, TX, VA.
Recall number
Z-2474-2021
Recall status
Terminated
CDRH record ID
188963
FDA record created
2021-09-14
Recall event number
88546

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗