Open original FDA record ↗Recall record
- FDA identifier
- Z-2475-2019
- Company in FDA record
- Alere San Diego, Inc.
- FDA record creation date
- Sep 5, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The polarity of the output of power supply was reversed. Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable.
Source facts
- Action
- On 2/15/2019, the firm provided a urgent medical device recall letter to all consignees who received the affected product. The letter was provided in a combination of methods to those consignees via direct mail, email or fax. The letter informed consignees of the following: 1. Alere San Diego, Inc. has concurrently sent a replacement Alere Cholestech LDX Power Supply at no extra cost to your shipping address on record via FedEx. Please refer to Purchase Order Number ASD119LDX for the replacement shipment. 2. " Discontinue use of and discard the affected Alere Cholestech LDX Power Supply. 3. If you have forwarded the product listed above to another facility, please provide a copy of this letter to them. 4. Complete and FAX or email the enclosed Verification Form within 10 days to confirm your receipt of this notice. 5. If you need to report any concerns you may have about the performance of the Alere Cholestech LDX System, please contact Technical Services at 1-877-308-8289 or by email at LDX.Support@alere.com. If you have questions, please contact Alere San Diego, Inc at 1-877-308-8289
- Initiated
- 2019-02-15
- Root cause
- Employee error
- Terminated
- 2023-12-12
- Distribution
- US Nationwide distribution in the states of AK, A, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY< OH, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI. No OUS consignees
- Recall number
- Z-2475-2019
- Recall status
- Terminated
- CDRH record ID
- 175109
- FDA record created
- 2019-09-05
- Recall event number
- 83410