Skip to content

Public FDA record

Z-2475-2023: LIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems,

Recall record Z-2475-2023 for Alphatec Spine, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2475-2023
Company in FDA record
Alphatec Spine, Inc.
FDA classification date
Aug 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.

Source facts

Action
On 7/7/23, recall notices were emailed to consignees who were asked to do the following: 1) Abstain from sale and/or use and return to the recalling firm. 2) Complete and return the response form via email to QMSCompliance@atecspine.com Questions/concerns can be directed to 1-800-922-1356 or QMSCompliance@atecspine.com
Initiated
2023-07-07
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of CT, IN, TX, LA, OH, AL, NH, MA, NC, AZ, MD, NJ, IL, CA, UT, MI.
Recall number
Z-2475-2023
Recall status
Open, Classified
CDRH record ID
202524
FDA classified
2023-08-28
Recall event number
92776

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗