- FDA identifier
- Z-2475-2023
- Company in FDA record
- Alphatec Spine, Inc.
- FDA classification date
- Aug 28, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.
Source facts
- Action
- On 7/7/23, recall notices were emailed to consignees who were asked to do the following: 1) Abstain from sale and/or use and return to the recalling firm. 2) Complete and return the response form via email to QMSCompliance@atecspine.com Questions/concerns can be directed to 1-800-922-1356 or QMSCompliance@atecspine.com
- Initiated
- 2023-07-07
- Root cause
- Nonconforming Material/Component
- Distribution
- US Nationwide distribution in the states of CT, IN, TX, LA, OH, AL, NH, MA, NC, AZ, MD, NJ, IL, CA, UT, MI.
- Recall number
- Z-2475-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202524
- FDA classified
- 2023-08-28
- Recall event number
- 92776