Open original FDA record ↗Recall record
- FDA identifier
- Z-2476-2020
- Company in FDA record
- Verathon, Inc.
- FDA classification date
- Jul 10, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which may result in video synchronization interruption, loss of live image/blank screen, no cameras connected monitor warning, may present risk to patients, in which time to intubation is more critical.
Source facts
- Action
- On 06/05/20, Field Safety Notice/ URGENT:MEDICAL DEVICE RECALL notices were sent to customers. Customers were asked to do the following: - Review the information contained within this Field Safety Notice with all staff members who should be aware of the contents of the Field Safety Notice. - Discontinue use of all affected cables. - When replacements are received, destroy or return all affected cables in your possession. - Complete and return the recall response form. Customers with additional questions are encouraged to contact Customer Care agent support Monday through Friday from 6:00am to 4:00pm Pacific Time for the United States at (800) 331-2313 or email CSNotifications@verathon.com
- Initiated
- 2020-06-05
- Root cause
- Device Design
- Terminated
- 2024-04-22
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Australia and United Kingdom.
- Recall number
- Z-2476-2020
- Recall status
- Terminated
- CDRH record ID
- 181588
- FDA classified
- 2020-07-10
- Recall event number
- 85705