Z-2476-2020: GlideScope Core OneTouch Smart Cable, REF: 0800 0601, an accessory to the GlideScope Video Laryngoscope S

Recall record Z-2476-2020 for Verathon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2476-2020
Company in FDA record
Verathon, Inc.
FDA classification date
Jul 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which may result in video synchronization interruption, loss of live image/blank screen, no cameras connected monitor warning, may present risk to patients, in which time to intubation is more critical.

Source facts

Action
On 06/05/20, Field Safety Notice/ URGENT:MEDICAL DEVICE RECALL notices were sent to customers. Customers were asked to do the following: - Review the information contained within this Field Safety Notice with all staff members who should be aware of the contents of the Field Safety Notice. - Discontinue use of all affected cables. - When replacements are received, destroy or return all affected cables in your possession. - Complete and return the recall response form. Customers with additional questions are encouraged to contact Customer Care agent support Monday through Friday from 6:00am to 4:00pm Pacific Time for the United States at (800) 331-2313 or email CSNotifications@verathon.com
Initiated
2020-06-05
Root cause
Device Design
Terminated
2024-04-22
Distribution
Worldwide Distribution: US (nationwide) and countries of: Australia and United Kingdom.
Recall number
Z-2476-2020
Recall status
Terminated
CDRH record ID
181588
FDA classified
2020-07-10
Recall event number
85705

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗