Open original FDA record ↗Recall record
- FDA identifier
- Z-2476-2021
- Company in FDA record
- Philips North America Llc
- FDA record creation date
- Sep 14, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
Source facts
- Action
- Philips notified accounts by Urgent Medical Device Recall letter via certified mailing United States Postal Service (USPS) on 8/13/21. Letter states reason for recall, health risk and action to take: Please place this Medical Device Recall letter with the documentation of the system until this correction has been implemented in your system. Additionally, please return the attached reply form to Phhilps to confirm that the users of the system have reviewed and understood this Medical Device Recall. Philips will correct the affected systems by replacing the ADUs. A Philips service representative will contact you to schedule the correct ion of your system. Please be assured that maintaining a high level of safety and quality is our highest priority. lf you need any further information or support concerning thls issue, please contact your local Philips representative: 1-800-722-9377 (reference FCO72200483
- Initiated
- 2021-08-13
- Root cause
- Device Design
- Terminated
- 2024-03-11
- Distribution
- US Nationwide distribution in the states of ID, MA, MI, TX, VA.
- Recall number
- Z-2476-2021
- Recall status
- Terminated
- CDRH record ID
- 188965
- FDA record created
- 2021-09-14
- Recall event number
- 88546