Skip to content

Public FDA record

Z-2476-2025: STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with th

Recall record Z-2476-2025 for Diagnostica Stago, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2476-2025
Company in FDA record
Diagnostica Stago, Inc.
FDA classification date
Aug 28, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.

Source facts

Action
Diagnostica Stago notified consignees on about 07/21/2025 via emailed letter titled URGENT: MEDICAL DEVICE RECALL (CORRECTION). Consignees were instructed to temporarily systematically run a quality control just before a batch of UFH or LMWH, if you run STA-LIQUID ANTI-Xa in UFH and LMWH applications on your STA SATELLITE analyzer. This request is being asked while Stago works to develop a permanent solution for the analyzer. Consignees were also requested to complete and return the Acknowledgement Form and to notify customers if affected units were transferred or further distributed.
Initiated
2025-07-21
Root cause
Mixed-up of materials/components
Distribution
Worldwide - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, LA, MD, ME, MN, MO, MT, NC, OH, OK, OR, PA, PR, SD, VT, WI, WY and the countries of Algeria, Argentina, Austria, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, France, French Polynesie, Guadeloupe, India, Iran, Israel, Jordan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Oman, Paraguay, Peru, Republique Tcheque, Romania, Saudi Arabia, Serbia, South Africa, Switzerland, Taiwan, Thailand, Tunisia, Turquie, United Arab Emirates, Uruguay, Vietnam.
Recall number
Z-2476-2025
Recall status
Open, Classified
CDRH record ID
215098
FDA classified
2025-08-28
Recall event number
97287

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.