Open original FDA record ↗Recall record
- FDA identifier
- Z-2477-2019
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA record creation date
- Sep 5, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Quality issue with high pressure solenoid valves
Source facts
- Action
- 1. For the cobas 8000 ISE, cobas c 701, and cobas c 702 modules and the cobas c 513 analyzer, stop using these units immediately. Your FES will be in contact with you to replace these affected high pressure solenoid valves immediately. 2. In the event that the high pressure solenoid valves cannot be replaced immediately, for the cobas e 601 module, follow the wash station checks outlined in this Urgent Medical Device Correction (UMDC). Perform these checks once per shift. If these checks show favorable outcomes and quality control results are within range, patient results may be reported. If these checks produce unfavorable outcomes, stop using your cobas e 601 module and call the Roche Support Network Customer Support Center at 1-800-428-2336. 3. Complete the attached fax form and fax or email it according to the instructions on the form. 4. File this Urgent Medical Device Correction (UMDC) for future reference Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
- Initiated
- 2019-07-18
- Root cause
- Device Design
- Terminated
- 2020-06-15
- Distribution
- State NY MT HI IN MD TX OH KS KY CA FL OR LA MO AK
- Recall number
- Z-2477-2019
- Recall status
- Terminated
- CDRH record ID
- 175616
- FDA record created
- 2019-09-05
- Recall event number
- 83525