Skip to content

Public FDA record

Z-2477-2023: epoc BGEM Crea Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1

Recall record Z-2477-2023 for Siemens Healthcare Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2477-2023
Company in FDA record
Siemens Healthcare Diagnostics Inc
FDA classification date
Aug 28, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.

Source facts

Action
Consignees were notified via an Urgent Medical Device Correction letter beginning June 5, 2023. Consignees were informed of the issue and were provided a workaround and actions to take. The next sensor configuration/software version, sensor configuration 42.1/epoc Host SW v3.38.2 /epoc NXS SW v4.11.11, is available to customers in Siemens Healthineers Document Library or the epoc Live Update Service (eLUS). This next sensor configuration, 42.1, mitigates the issue.
Initiated
2023-06-01
Root cause
Under Investigation by firm
Distribution
US distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam
Recall number
Z-2477-2023
Recall status
Open, Classified
CDRH record ID
201323
FDA classified
2023-08-28
Recall event number
92537

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.