Skip to content

Public FDA record

Z-2477-2025: BD Pyxis Product Name / UDI-DI code / Catalog No. BD Pyxis" MedStation" 4000 Main / 10885403512629 / 303

Recall record Z-2477-2025 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2477-2025
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Aug 28, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Antivirus software was not consistently installed on impacted devices during the implementation process.

Source facts

Action
On 05/08/2025, the firm sent via mail and email an "URGENT MEDICAL DEVICE PRODUCT CORRECTION" Letter informing customers of two issues. For this recall the issue pertains to technicians failing to install (ESET) antivirus software on affected devices. This can potentially result in a malicious actor that could exploit a cybersecurity weakness on a device that is configured without antivirus protection within a hospital network. Customer are instructed and/or informed: For Clinical Users: -to utilize user facility policies and procedures for cybersecurity issues or disaster recovery. -to remove devices from network and continue to use operational devices in critical override per the BD Pyxis" MedStation" ES user guide DME 10000430667, version 00 (pages 74-75). For IT, Safety and Security: -Firewall Protection - The hospital firewall can block vulnerable ports. -Limit physical access to only authorized personnel. -Tightly control management of system passwords provided to authorized users. -Monitor and log network traffic attempting to reach the affected products for suspicious activity. -Isolate affected products in a secure VLAN or behind firewalls with restricted access that only permits communication with trusted hosts in other networks when needed. -Physical security - Manage who can access network or server rooms to reduce cyberattack or data breach risks. -Secure user access implement secure user passwords and/or biometric logins at the application level. -Regular database backups (ES devices only). -Additional Security layers - There are other layers of security still in place to mitigate potential malicious activity on ES devices such as secure Operating System passwords, database and file encryption, and secure user passwords / biometric logins at the application level. For questions: BD Remediation Support Phone: 1-866-583-8783 Phone Hours: 5:00am PST to 4:00pm PST (8:00am ET to 7:00pm ET) Monday - Friday Recall Customer Response Forms - email: BDRC22
Initiated
2025-05-08
Root cause
Under Investigation by firm
Distribution
Worldwide - U.S Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of United Arab Emirates, Argentina, Australia, Belgium, Bahrain, Bermuda, Brazil, Bahamas, Canada, Switzerland, Germany, Egypt, Finland, France, United Kingdom, Hong Kong, Indonesia, Ireland, India, Italy, Jordan, Lebanon, Morocco, Mexico, Malaysia, New Zealand, Qatar, Saudi Arabia, Singapore, Thailand, Greece, Oman, and Taiwan.
Recall number
Z-2477-2025
Recall status
Open, Classified
CDRH record ID
213001
FDA classified
2025-08-28
Recall event number
96477

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗