Z-2478-2021: Medtronic HVAD Pump Implant Kits

Recall record Z-2478-2021 for Heartware, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2478-2021
Company in FDA record
Heartware, Inc.
FDA record creation date
Sep 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is more force required to slide the driveline cover (boot cover) away from the connector after completion of a Driveline Strain Relief Repair than over an unrepaired strain relief. This may lead to difficulty in pulling back the driveline cover when attempting to access the Driveline connector during a controller exchange. Additionally, sliding the driveline cover over a strain relief repair may compromise the repair.

Source facts

Action
Beginning 10-Aug-2021 (US), and no later than 17-Aug-2021 (OUS), Medtronic initiated a Global notification to inform HVAD clinicians (following physicians and VAD Coordinators at HVAD accounts) who are currently treating patients implanted with the HVAD system about the potential for difficulty in pulling back the HVAD Pump Driveline cover for drivelines that have received a strain relief repair. In the United States, beginning 10-Aug-2021, an Urgent Medical Device Notice, was delivered to following physicians and VAD Coordinators, who are currently caring for an HVAD patient, via UPS 2-day delivery as well as through verbal and in person follow ups from the Medtronic field staff. Also, in the United States, on 17-Aug-2021, Medtronic sent a notification to active HVAD Patients in the U.S. according to their registered contact information via UPS 2-day delivery to make them aware of the potential for this issue and to reinforce precautions in the IFU and Patient Manual cautioning against selfrepairing the HVAD equipment. Outside of the United States, beginning 10-Aug-2021 and no later than 17-Aug-2021, a formal FCA notification will be delivered to VAD Coordinators and following physicians who according to Medtronic records are treating patients implanted with an HVAD system. This letter will be delivered using locally approved methods. Confirmation of the communication will be received via locally approved methods.
Initiated
2021-08-06
Root cause
Device Design
Distribution
Distribution throughout the US. Also distribution in ARGENTINA Armenia, Australia, Austria, BELGIUM, BRAZIL, CANADA, CHILE, Colombia, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, Germany, GREECE HONG KONG, HUNGARY, INDIA, ISRAEL, Italy, JAPAN, Kazakhstan, Korea, KUWAIT, Latvia, LEBANON, LITHUANIA, LUXEMBOURG, Macedonia, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, Poland, ROMANIA, SAUDI ARABIA, Serbia, Singapore, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, Taiwan, TURKEY, Ukraine, United Arab Emirates, United Kingdom, VIETNAM
Recall number
Z-2478-2021
Recall status
Open, Classified
CDRH record ID
188956
FDA record created
2021-09-16
Recall event number
88541

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗