Open original FDA record ↗Recall record
- FDA identifier
- Z-2479-2018
- Company in FDA record
- Hitachi Medical Systems America Inc
- FDA record creation date
- Jul 18, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet for scanning. The firm has received two complaints where technologists have cut their fingers on this belt.
Source facts
- Action
- On June 18, 2018, the firm distributed an Urgent Medical Device Correction Notices to affected customers. Customers were informed of the issue and advised that the firm would be adding protective guards to both sides of the belt to prevent accidental contact. The firm expects to begin installing the guards around fall 2018. In the meantime, customers are advised to avoid direct contact with the stainless steel belt. Questions, comments, or requests may be sent to recalls@hitachihealthcare.com.
- Initiated
- 2018-06-18
- Root cause
- Device Design
- Terminated
- 2020-04-20
- Distribution
- Worldwide Distribution - US Nationwide distribution.in the states of Mexico.
- Recall number
- Z-2479-2018
- Recall status
- Terminated
- CDRH record ID
- 165552
- FDA record created
- 2018-07-18
- Recall event number
- 80464