Z-2479-2019: OMNIlife science, Inc. ApeX acetabular liner hip prosthesis component Serf Dual Mobility Cup Liner 53-

Recall record Z-2479-2019 for OMNIlife science Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2479-2019
Company in FDA record
OMNIlife science Inc.
FDA record creation date
Sep 5, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Acetabular liners failed the oxidation testing performed prior to release. They may experience a higher wear rate, leading to revision surgery.

Source facts

Action
OMNIlife science, Inc. sent an Urgent Medical Device Recall letter, dated July 23rd, 2019. 1. Identify the affected product from inventory and hold (quarantine) the product. 2. Return product to Omni at 480 paramount drive, Raynham, MA 02767 3. Acknowledge this notice by retuning the acknowledgement and receipt form by fax to 508-819-3390 or email to Omni-cs@coringroup.com For further information contact Christina Rovaldi, Regulatory Affairs Manager at 774-226-1847 or 800-448-6664. Return the attached Acknowledgement form by FAX to 508-819-3390 or by email to Omnics@coringroup.com. Company contact hours are Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time.
Initiated
2019-07-23
Root cause
Nonconforming Material/Component
Terminated
2020-04-24
Distribution
US nationwide distribution to states of: IL, NY, OK, MD, AL, PA, CO, FL, and UT.
Recall number
Z-2479-2019
Recall status
Terminated
CDRH record ID
175623
FDA record created
2019-09-05
Recall event number
83528

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗