Open original FDA record ↗Recall record
- FDA identifier
- Z-2479-2023
- Company in FDA record
- Edwards Lifesciences, LLC
- FDA classification date
- Aug 29, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance when attempting to release the implant from the delivery system by rotating the release knob. Additional torque required to initiate rotation of the release wire from the implant, could cause implant rotation or additional implant implantation.
Source facts
- Action
- On 7/27/23, recall notices were sent to customers and distributors who were asked to do the following: 1) Stop using and work with your firm representative to return affected devices. 2) Share this notice with the appropriate clinical staff at your site. 3) Complete and return the customer acknowledgement form. 4) Distributors: Notify your customers by sending this customer notification to any of your customers who have purchased the impacted devices.
- Initiated
- 2023-07-27
- Root cause
- Process change control
- Distribution
- International distribution to the countries of Austria, Belgium, Switzerland, Germany, Cyprus, Israel, UK, Greece, Ireland, Italy, Luxembourg, Poland.
- Recall number
- Z-2479-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202625
- FDA classified
- 2023-08-29
- Recall event number
- 92804