Open original FDA record ↗Recall record
- FDA identifier
- Z-2481-2015
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Aug 28, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Beckman Coulter is recalling the LH 750 and LH780 Hematology Systems because the use of the diluent lots may eventually result in a compromised Hgb cuvette.
Source facts
- Action
- An Urgent Medical Device Recall letter dated 7/22/15 was sent to the affected customers to inform them of the issue and actions to take. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions are instructed to contact Beckman Coulter's Customer Support Center, http://www.beckmancoulter.com/customersupport/support, (800) 526-7694 in US and Canada. Customers outside of the US and Canada, are instructed to contact their local Beckman Coulter Representative.
- Initiated
- 2015-07-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-12-18
- Distribution
- Distributed US (nationwide) including Puerto Rico and U.S. Virgin Islands and the countries of Bermuda, Canada, Guatemala, Hong Kong, Mexico, Panama, Taiwan, and Trinidad-Tobago.
- Recall number
- Z-2481-2015
- Recall status
- Terminated
- CDRH record ID
- 139181
- FDA classified
- 2015-08-28
- Recall event number
- 71835