Open original FDA record ↗Recall record
- FDA identifier
- Z-2482-2019
- Company in FDA record
- Haag-Streit USA Inc
- FDA record creation date
- Sep 5, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
software error in the central control unit of the floor stand.
Source facts
- Action
- 1. Do not perform any other functions at the same time when operation the XY control, such as changes in the light intensity. 2. Perform a software update of the floor stand. you will receive the software version 11.4 through your local representative which will fix the bug. 3. please send us the enclosed response form filled In as soon as possible. 4. If you have passed the products to third parties, please forward a copy of this information and inform the below mentioned contact person. The software update will be organized through Haag-Streit USA. For questions concerning this procedure, please contact: Brian Gallagher Phone: (513) 486-1522 Email: fieldsafetynotice@haag-streit-usa .com
- Initiated
- 2019-07-31
- Root cause
- Software Manufacturing/Software Deployment
- Terminated
- 2020-04-16
- Distribution
- US in the state of OH
- Recall number
- Z-2482-2019
- Recall status
- Terminated
- CDRH record ID
- 175754
- FDA record created
- 2019-09-05
- Recall event number
- 83571