Open original FDA record ↗Recall record
- FDA identifier
- Z-2482-2021
- Company in FDA record
- Welch Allyn Inc
- FDA record creation date
- Sep 16, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If the device is exposed to fluid ingress and is used before the cleaning fluid has completely dried, there is a risk of the device overheating potentially causing a burn to the user or the ear canal of the patient
Source facts
- Action
- Welch Allen (a Hill Rom company) issued Urgent Medical Device Correction Letter (FA-2021-05-001-SKF-004) on 7/30/21 to: Chief Executive; Facility Administrator; Facility Engineer; Vigilance Manager; Biomedical Engineering; Medical Device Liaison Officer; Distributor mailed (with signature service) to all first-level consignees by Sedgwick (third party recall company). The letter states reason for the Field Action, risk mitigations, and Actions to be taken by the User: 1. Please share this communication with all potential users in your organization and instruct them to follow the provided Cleaning Guide for proper cleaning. 2. Do not use the device if the ring around the measurement button shows a green blinking or flashing light instead of a ready state (solid green light) and contact Hillrom Customer Service to report this issue. 3. Do not use the device if the device requires multiple power-ups prior to going to ready state (solid green light) and contact Hillrom Customer Service to report this issue. 4. If you experience an overheating probe tip, do not use the device and contact Hillrom Customer Service to report the issue. 5. Complete the attached response form and return to HillromSKF004@stericycle.com within one month. Actions to be taken by the distributor: Please share with end users and/or your accounts and complete the attached response form and return to HillromSKF004@stericycle.com within one month. Contact HillromSKF004@stericycle.com to receive an electronic copy of this notification, response form, and further instructions for notifying your accounts. Actions being taken by Hillrom: Hillrom is in the process of updating the Braun ThermoScan Pro 6000 Instructions for Use (IFU) to include additional fluid ingress warnings. The IFU and Cleaning Guide can be found on the Hillrom website, Hillrom.com, under the Education & Documentation section of the Braun ThermoScan PRO 6000 Products page. The updated IFU will be available on the website
- Initiated
- 2021-07-20
- Root cause
- Labeling design
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Afghanistan, Andorra, Australia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Canada, Chile, China, Colombia, Congo, Costa Rica, C¿te D'ivoire (Ivory Coast), Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep., Ecuador, Estonia, Finland, France, Frenc.Polynesia, French, Guiana, Germany, Greece, Guadeloupe, Guam, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kenya, Kuwait, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Malta, Martinique, Mexico, Monaco, Myanmar, N.Mariana Islnd, Netherlands (the), New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion Romania, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Ukraine, United Kingdom, Utd.Arab Emir., Vietnam, Zimbabwe.
- Recall number
- Z-2482-2021
- Recall status
- Open, Classified
- CDRH record ID
- 188768
- FDA record created
- 2021-09-16
- Recall event number
- 88481