Z-2483-2021: EPIQ Diagnostic Ultrasound Systems: Software Versions 5.0, 5.0.1 and 5.0.2 Models: EPIQ 5 - 795120 EPIQ

Recall record Z-2483-2021 for Philips Ultrasound Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2483-2021
Company in FDA record
Philips Ultrasound Inc
FDA record creation date
Sep 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software defect that can intermittently cause the system to lock-up which exiting Review Mode while performing an exam.

Source facts

Action
On 08/06/2021, the firm sent via certified mail an "URGENT Medical Device Correction" Notification Letter to customer informing them that due to a software defect, when reviewing or adjusting xPlane exam results, there is a potential for the ultrasound systems to become unresponsive (Lock-up) preventing users from continuing clinical use. The ultrasound system provides an error notification including method to restart and bring system back to normal use. The Recalling Firm has received a report of an ultrasound system locking up multiple times during an open cardiac procedure while using a Transesophageal Echocardiogram (TEE) transducer. Each lock-up required at least one system reboot to regain functionality. Customers are instructed to: 1) Determine the software version on their ultrasound system by: - Power up the system and allow it to complete the boot sequence; - Press Support on the right side of the control panel; - Under System Management, click System Information; - The software version is listed in the Software Information Section. 2) Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm. 3) Post this notice near the affected unit(s) for ease of reference. 4) Complete and return the response form to the Recalling Firm promptly and no later than 30 days from receipt via email to: ultrasound.corrections@philips.com or fax to 1-833-512-7756. In addition, because the software defect may intermittently cause the system to lock-up when exiting Review mode, customer are asked to take the following steps to minimize the likelihood of occurrence: 1. Acquire a loop while in xPlane, xPlane Doppler, or Dual mode. 2. Do NOT adjust an imaging control (for example: Rotate, Tilt, Gain) while the loop capture is in progress.* (*If the imaging controls have been adjusted during the loop capture, Philips recommends that the system be rebooted prior to reviewing the loop in Review mode). 3. Allow the loop
Initiated
2021-08-06
Root cause
Software design
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Albania, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Brazil, Brunei Darussalam, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, French Guiana, Germany, Gibraltar, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Virgin Island, Yemen and Zambia.
Recall number
Z-2483-2021
Recall status
Completed
CDRH record ID
188886
FDA record created
2021-09-17
Recall event number
88514

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗