Z-2483-2023: Oculus Pentacam AXL , Ref 70100, CE 0123

Recall record Z-2483-2023 for Oculus Optikgeraete GMBH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2483-2023
Company in FDA record
Oculus Optikgeraete GMBH
FDA classification date
Aug 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is a potential that optical devices with insufficient anti-reflective coating may lead to incorrect axial length measurements.

Source facts

Action
On August 04, 2023, Oculus issued a Urgent Field Safety Notice" Correction via UPS Express. Oculus ask consignees to take the following actions: 1. Immediately stop using the affected devices for axis length measurement (mark the affected devices with a note) or take them out of service immediately. Devices not listed in Appendix B are not affected by this corrective action and can continue to be used according to the instructions for use. 2. Please fill out the attached form (Appendix A) and return it to OCULUS and your local OCULUS representative as soon as possible, but no later than 30 days after receipt. By completing this form, you acknowledge receipt of the Urgent Field Safety Notice and that you understand the issue and the actions required. 3. Pass this urgent safety information on to all users of the affected products and other people who need to be informed so that they are aware of the problem. Please file this letter directly with your device and make sure that it is and remains visible to all users at the storage location. It is important that the meaning of this notification is understood. 4. If you have given the products to a third party, please forward a copy of this information immediately and please inform us accordingly. 5. Please keep this urgent safety information at least until the action has been completed.
Initiated
2023-08-08
Root cause
Process control
Distribution
US Nationwide distribution in the states of CA, FL, MA, MN, NY, NC, OH, PA, TX, VA, WA. OUS: Not Provided
Recall number
Z-2483-2023
Recall status
Open, Classified
CDRH record ID
202699
FDA classified
2023-08-30
Recall event number
92847

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗